Crucially, Narishige designed the PC-100 to be . This means that many parts, such as the heating element (PC-10H) and glass capillaries, are the same for both models, ensuring that existing users are not left without critical consumables.
The glass capillary must sit perfectly centered inside the platinum loop. If the capillary leans closer to one side of the coil, one side of the glass will melt faster, resulting in bent or hooked pipette tips. Loosen the coil mounting screws to manually center the loop if necessary. Troubleshooting Matrix Probable Cause Corrective Action Off-center heating coil or uneven airflow. Center the coil; shield the unit from drafts. Glass breaks instantly Heater setting is too low; weight is too heavy. Increase the heater dial value incrementally. No heat generated Burned-out platinum coil or blown fuse. Inspect the coil for breaks; check the rear fuse. Inconsistent tip sizes Variations in glass OD or sliding friction. Clean the vertical guide rods with a dry cloth. narishige pc10 manual new
: Heating and electromagnetic values are displayed as digital units (0–100), where a heater level of 100 corresponds to approximately 2.5V. This allows for reproducible mass production. Crucially, Narishige designed the PC-100 to be
Yes. Narishige originally published an English version (Part No. 3304-01E). The most common "new" PDF circulating is a 24-page document dated 1988. Verify it includes the calibration flowchart. If the capillary leans closer to one side
Securely plug in the AC cord; check and replace fuses in the inlet. Internal controller malfunction. Contact Narishige Service. Uneven Pulls Dirty filament or misaligned glass.
Several academic institutions and biophysics forums host "abandoned manual" libraries.
The PC-10 manual dictates a specific sequence for creating a pipette 1.2.1. Step 1: Single Stage Pulling (For Preliminary Tapering) Set the Mode-selector knob to STEP 1 . Adjust the NO. 1 HEATER level to the desired value. Press the Start switch. Step 2: Two-Stage Pulling (For Finalizing Tips) For a two-stage pull (essential for finer tips): Set the mode to STEP 2 . Set the NO. 1 HEATER level.